A centrifuge is a simple machine that fails in expensive ways. These five checks take under two minutes and prevent almost every bench failure we see reported.
1. Balance the load. Opposite positions must carry equal masses within one gram. Water-fill blanks if you have an odd number of samples.
2. Seat every tube. Tubes must sit flat on the bottom of the bucket or fixed-angle rotor. A tube on an angle is a projectile waiting for its cue.
3. Inspect tubes for cracks. Especially glass tubes after autoclaving. A cracked tube in a rotor can shatter the rotor.
4. Clean the rotor chamber. Blood, saline and reagent spills corrode aluminium rotors. Wipe after every run.
5. Confirm the lid latch. Do not attempt to start until the interlock has clicked closed.